Model
CNSolution 9310 Analyzer
Made by Ezkem Corporation · Hood River, OR
- Ni
- 0-1 mg/L
- Scn
- 0-500 mg/L
- So2
- 0-200 mg/L
- Tds
- 8860 mg/L
- Duty
- continuous 24 h/365 d
- Ocn-
- 196 mg/L
- S2o3
- 0-200 mg/L
- S4o6
- 0-100 mg/L
- Scn-
- 206 mg/L
- Size
- 48.3 cm H x 31.1 cm W x 31.1 cm D
- Nh3-n
- 13.0 mg/L
- No2-n
- 0-10 mg/L
- No3-n
- 0-10 mg/L
- Power
- 24VDC
- Range
- 0.2‑50 ppm CN, 2‑500 ppm CN, 20‑2000 ppm CN
- Weight
- 11kg
- Accuracy
- ±5%@50ppm
- Function
- available cyanide measurement in leach tanks
- Enclosure
- NEMA4X
- Sample ph
- 9‑11
- Technique
- gas‑diffusion amperometry
- Dimensions
- 48.3x31.1x31.1cm
- Calibration
- 2‑point
- Reagent naoh
- 0.1 M NaOH, 10 L/week
- Analysis time
- <3 min
- Reagent water
- CN‑free water, 10 L/week
- Sample volume
- <1 mL
- Detection limit
- 0.05 mg/L CN
- Measurement ranges
- 0.2‑50,2.0‑500,20‑2000ppm
- Reference methods
- OIA‑1677, ASTM D 6999‑09
- Product class
- Analytical and scientific instruments, excluding opticalNAPCS 2017725000
From Ezkem Corporation’s own documents · read October 12, 2026 · not verified by FactoryRegistry
Documents
Where this comes from — 4 documents
A machine read this from documents Ezkem Corporation publishes on its own website, as of October 12, 2026. FactoryRegistry has not verified it; the documents are the authority.
- 4036spec.fm, page 1: “function: available cyanide measurement in leach tanks”
- The CNSolution™ 9310 On-line Cyanide Analyzer, page 20: “no3-n: 9.7 mg/L”
- CNSolution9310, page 5: “Operating Principle FIA by gas diffusion amperometry”
- CNSolutionTM9310: “CNSolutionTM9310”
- CNSolutionTM9310, page 3: “Measurement Ranges 0.2 to 50 / 2.0 to 500 / 20 to 2000 ppm CN; Accuracy ± 5 % at 50‑ppm; Power Requirements 24 VDC; External Dimensions 48.3 cm H x 31.1 cm W x 31.1 cm D; Weight 11 kg; Instrument Enclosure NEMA 4X / IEC …”